720 series II pH/ORP MONITOR/CONTROLLERS
Click here for detailed image of the 720 pH
Click here for detailed image of the 720 ORP

 

BENEFITS

  • Affordable

  • Ease of Operation

  • Low Maintenance

  • Ensures Product Quality

The advanced “isolated” circuitry of the 720 Series II pH/ORP Monitor/ controllers guarantees accurate and reliable measurements — completely eliminating ground-loop and noise issues.

 

The unique sensor preamp allows for longer distances between the sensor and the Monitor/controller without the loss of accuracy or reliability.

 

All Myron L Monitor/controllers feature a highly refined and precise Temperature Compensation circuit. This feature perfectly matches the NERNST equation correcting the displayed reading to 25°C. The TC may be disabled to conform with USP requirements.

 

Built-in electronic calibration allows for fast quality checks without buffers. (Note: for maximum system accuracy buffers are recommended).

 

A push-to-test button makes set point checks easy, while the internal alarm set discourages unauthorized adjustments.

 

Standard on all controller models is a heavy-duty 10 amp output relay, operating on either increasing or decreasing readings. Bright green/red LEDs indicate above/below set point readings.

 

All 720 Series II models feature an IP65/ NEMA 4X water resistant and corrosion proof rated enclosure. Compact in size, at 152 x 122 mm/6 x 4.8 in., the 720 Series II models are suitable for panel, bench or surface mounting.

 

Additional options include a Remote Alarm™ model RA.

 

Applications

Process Control
Reverse Osmosis
Seawater Desalination
Waste Treatment
Food Processing
Plating
Power Plants
Laboratories
Printing
Paper & Pulp
Boilers & Cooling Towers
Potable Water
Waste Water Treatment
Swimming Pools & Spas
Bleach Manufacturing
Reduction of Chromate Waste
Environmental
 
 

FEATURES

pH 0-14 or ORP ± 2000 mV
Choice of 3˝ digit display or analog meter
Advanced "Isolated" circuitry
Automatic temperature compensation to 25°C (NERNST equation)
Isolated 0-10 VDC Recorder Output standard
IP65/NEMA 4X water resistant and corrosion-proof enclosure
Built-in electronic calibration
Single set point/alarm standard
Adjustable set point hysteresis 
Solid State 24 VDC Output
10 amp relay operates on choice of increasing or decreasing reading
Push-to-test set point check
Internal set point discourages unauthorized adjustment
Selectable 115/230 VAC
Unique, easily installed sensors
CE Certified
Unique, easily installed sensors
Heavy duty “F” model sensors may be installed in ANY direction including inverted

 

pH/ORP SENSORS

 

 

These highly accurate sensors utilize a preamplifier to ensure stability over longer distances, and to eliminate potential noise problems thereby ensuring accurate repeatable readings.

 

In addition, these reliable sensors feature an integral temperature sensor to ensure accurate and rapid automatic temperature compensation.

 

Several choices of sensors are available, in 1/2” MNPT or 3/4” MNPT, depending on the demands of the application. The basic pH sensor utilizes the standard round glass electrode with a single reference junction. The basic ORP sensor utilizes a Platinum electrode with a single reference junction. See 720II Sensor Selection Guide for additional information.

 

NOTE:  Heavy Duty “F” model sensors may be installed in ANY direction including inverted.

 

 

SENSOR SPECIFICATIONS

Temperature Compensation: Integral temperature sensor provides automatic TC to 25°C between 0 - 100°C/ 32 - 212°F.

Pressure/Temperature Limits: P72S: Intermittent use up to 50 psi/690 kPa 100°C/212°F. High Temperature models available, see pH/ORP Sensor Selection Guide.
Cable: 5 wire, 24 ga. shielded, 10 ft./3 meters standard. 25 and 100 ft./8 and 30 meter lengths available.
Material: 1/2" MNPT: CPVC body. 3/4" MNPT: Ryton®* body.

Dimensions: P72S typical: overall length 165 mm/6.5 in., body 127 mm/5 in., electrode length 19 mm/.75 in.

 

*Ryton® PPS is a registered trademark of Chevron Phillips Chemical Company LLC

 

 

Additional Information

720 Series II Data Sheet
(includes specs, options and order info)

click here: PDF

Price List click here: PDF
OEM Parts click here: PDF
Replacement Parts click here: PDF
720II Sensor Selection Guide  click here: PDF

Operation Manual

click here: PDF

 

Limited Warranty
All Myron L Monitor/controllers have a Two (2) Year Limited Warranty. pH Sensors have a Six (6) Month Limited Warranty. If any Monitor/controller fails to function normally, return the unit to the factory prepaid. If, in the opinion of the factory, failure was due to materials or workmanship, repair or replacement will be made without charge.