For more information, please review our PT3 Literature:
ULTRAPEN PT3 DATA SHEET
ORP | Temperature |
Range: | -1000 mV to +1000 mV | 0 - 71°C / 32 - 160°F |
Accuracy: | ± 10 mV | ± 0.1ºC / ± 0.1ºF |
Resolution: | 1 mV ORP | 0.1ºC / 0.1ºF |
Time to Reading Stabilization: 10 - 45 seconds | ||
Power Consumption: Active Mode 37 mA, Sleep Mode 2 µA | ||
Temperature Compensation: Automatic In Calibration Mode From 15ºC to 30ºC* | ||
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D | ||
Weight: 50.4 g / 1.78 oz. (without soaker cap and lanyard) | ||
Case Material: Anodized Aircraft Aluminum with Protective Coating | ||
Battery Type: N type, Alkaline, 1.5V | ||
Calibration Standard Solutions: ORP80, ORP260, ORP470 | ||
Operating/Storage Temperature: 0 - 55ºC or 32 - 131ºF | ||
Enclosure Ratings: IP67 and NEMA 6 | ||
EN61236-1: 2006 – Annex A: 2008: Electrostatic discharge to the PT3 may cause it to |
* Temperature compensation in calibration mode: Temperature affects the reaction potentials for all chemicals differently. True ORP is the direct measurement of electron activity during an oxidation-reduction reaction, regardless of temperature. However, for maximum accuracy and ease of calibration, the Myron L Company has developed three calibration standard solutions with known dissolved species. We derived the temperature compensation (from 15șC to 30șC) for those solutions, and embedded automatic temperature compensation into the calibration function of your PT3. Note: To verify calibration while in measurement mode, you must manually correct for any variation in temperature. Example: @25șC, ORP2602OZ calibration solution will read 260mV, however @ 20.0șC ORP2602OZ will read 265mV.
Standard Solutions for calibration:
The ULTRAPEN™ PT3 requires Myron L Company 80mV Quinhydrone, 260mV
Quinhydrone, or 470mV MLC Light’s ORP Standard Solutions for wet calibration and
Sensor Storage Solution for proper storage.
Order model#s: | |
ORP802OZ 80 mV | Kit - Comes with Quinhydrone powder and 2oz. bottle of PH7 solution |
ORP2602OZ 260 mV | Kit - Comes with Quinhydrone powder and 2oz. bottle of PH4 solution |
ORP4702OZ 470 mV | MLC Light's Solution - 2oz. bottle |
ORPCALKIT | ORP Calibration Kit - includes all required solutions and powders for 80mV, 260mV, and 470mV calibration solutions** |
**For additional calibration kits available, contact the Myron L Company | |
PHSS | Storage Solution - Available in 2 oz. or quart sized bottles |
Replacement ORP Sensor:
The ULTRAPEN™ PT3 features a user replaceable ORP sensor. Installation instructions included with sensor.
Order model#: RPT3
ORP Electrode Cleaning Paper:
Routinely cleaning the electrode will keep the sensor free of build-up that can affect the speed and accuracy of ORP readings. Instructions for use are included with the paper.
Order model#s: OECP-25 for QTY 25 or OECP-100 for QTY 100
The PT3, excluding the ORP sensor, has a one (1) year limited warranty.