We are proud of the trust our handheld water quality instruments and monitor/controllers have earned in the past. Our product line has evolved to a new level of outstanding performance and value in analytical instruments: the Ultrameter II series. While priced like affordable single-parameter instruments, the Ultrameter II does the job of up to seven instruments.
Both Ultrameter II models deliver performance of ±1% of reading (not merely full scale). This high level of accuracy has been achieved through advanced four-electrode conductivity cell technology, a unique pH/ORP sensor and powerful microprocessor-based circuitry. With displayed values of up to 9999, the full four-digit LCD ensures resolution levels never before possible in such affordable instruments. Factory calibrated with NIST traceable solutions, each Ultrameter II may be supplied with both certification of traceability and NIST traceable solutions for definitive calibration.
The Ultrameter II is a prime example of how high-tech engineering can greatly simplify and streamline a task. Whether in the lab, industrial plant, or in a remote field location, merely:
Temperature compensation and range selection are both rapid and automatic. The Ultrameter II is a true one-hand operation instrument.
All calibrations are quickly accomplished by pressing the ▲ or ▼ keys to agree with our NIST traceable Standard Solution. When calibration is necessary, display prompts simplify pH calibration and make sure the correct buffer is being used. Plus, all parameters (excluding factory-set temperature) have an internal electronic setting that can be used for field calibration and as a check on pH/ORP sensor life.
The Myron L® Company FCE function reports FAC quickly and accurately by measuring ORP, the chemical characteristic of chlorine that directly reflects its effectivity, cross referenced with pH. Both DPD kits and colorimeters may tell the user the FAC value of the sample in the test tube, but since the chemistry of that sample is quite different from the source water being analyzed, the results are imprecisely related to actual disinfection power. The Myron L® Company FCE function measures the real, unaltered chemistry of source water, including moment-to-moment changes in that chemistry.
Beyond their affordable purchase price, Ultra-Fast, Ultra-Easy, Ultra- Powerful Ultrameter IIs save both time and money. Measure for measure, Ultrameter IIs give you a better return on your investment than any other handheld instrument. To see for yourself, contact your distributor or the Myron L® Company today.
ULTRAMETER™ II |
Display: 4 Digit Liquid Crystal Display |
Dimensions: L x W x H 196 x 68 x 64 mm/7.7 x 2.7 x 2.5 inches |
Weight: 352 g/12.4 oz. |
Case/conductivity cell material: VALOX* |
Cell capacities pH/ORP: 1,2 ml/0.04 oz. Cond/TDS/Res: 5 ml/0.2 oz. |
Power: 9V alkaline battery |
Battery life: >100 hours (5000 readings) |
Operating/storage temperature: 0 – 55°C/32 – 132°F |
Protection Ratings: IP67/NEMA 6 (waterproof to 1 meter/3 feet) |
EMI/EMC Ratings (Conformité Européenne): IEC 61326-1: 2020; CISPR 11: 2015 +A1: 2016 +A2: 2019; IEC 61000-4-2: 2008; IEC 61000-4-3: 2010; IEC 61000-4-8: 2009; (EU) 2015/863 (RoHS3): +(EU) 2024/232 |
Conductivity | Ranges: 0–9999 μS/cm; 10–200 mS/cm; in 5 autoranges |
Resolution: 0.01(<100 μS); 0.1(<1000 μS); 1(<10 mS); 0.01(<100 mS); 0.1(<200 mS) | |
Accuracy: ±1% of reading | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Adjustable Temperature Compensation to 25°C: 0–9.99%/°C | |
Conductivity/TDS Ratios Preprogrammed: KCl, 442*, NaCl | |
Adjustable Conductivity/TDS Ratio Factor: 0.20–7.99 | |
TDS | Ranges: 0–9999 ppm; 10–200 ppt; in 5 autoranges |
Resolution: 0.01(<100 ppm); 0.1(<1000 ppm); 1(<10 ppt); 0.01(<100 ppt); 0.1(<200 ppt) | |
Accuracy: ±1% of reading | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Adjustable Temperature Compensation to 25°C: 0–9.99%/°C | |
Conductivity/TDS Ratios Preprogrammed: KCl, 442*, NaCl | |
Adjustable Conductivity/TDS Ratio Factor: 0.20–7.99 | |
Resistivity | Ranges: 10 KΩ–30 MΩ |
Resolution: 0.01(<100 KΩ); 0.1(<1000 KΩ); 0.1(>1 MΩ) | |
Accuracy: ±1% of reading | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Adjustable Temperature Compensation to 25°C: 0–9.99%/°C | |
pH | Ranges: 0–14 pH |
Resolution: 0.01 pH | |
Accuracy**: ±0.01 pH | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
ORP | Ranges: ±999 mV |
Resolution: 1 mV | |
Accuracy**: ±1 mV | |
Free Chlorine FCE™ | Ranges: 0.00–9.99 ppm when 350≤ORPmV<725 and 0.0≤pH<9.9 and when 725≤ORPmV<825 and 0.0≤pH<8.9 |
Resolution: 0.01 ppm | |
Accuracy**: <1.00ppm ±0.3ppm and ≥1.00ppm ±0.2ppm | |
Auto Temperature Compensation: 0–71°C / 32–160°F | |
Temperature | Ranges: 0–71°C / 32–160°F |
Resolution: 0.1°C/F | |
Accuracy: ±0.1°C |